Controller Specifications
Tab. 2
Input:
Voltage
Frequency
Power
Fuse
Output:
Voltage
Frequency
Power
Environment storage
Operating
Applied norr.
Weight:
Pollution Degree
Installation Category
Acoustic noise (ISO7779)
The controller supplied voltage must be easily disconnected (through a circuit
NOTE
bracker or having access to the power supply cable plug).
TwisTorr 74FS AG Rack Controller User Manual / 87-901-055-01
Technical Information
Controller Specifications
100 to 240 Vac (voltage fluctuation +/- 10%)
50 to 60 Hz
450 VA
2 x T4 A (slow blow) 250 V
75 Vac
1167 Hz
100 W (Ramp), 70 W (Normal and Autotuning)
-20 °C to +70 °C
0 – 95 %
+5 to +45 °C
0 – 90 %
EN 61010 -1
EN 55011 (Class-A)
EN 61000-3-2,
EN 61000-3-3,
EN 61000-6-2,
EN 61000-4-2,3,4,5,6,11, 8
1.7 Kg (3.2 lbs)
2
II
LpA < 70 dB(A), normal operation, operator position
17
211/300