Controller Specifications
Tab. 16
Input:
Voltage
Frequency
Power
Output for pump:
Power during pump ramp-up
Power normal (pump water cooled)
Power normal (pump air cooled)
Operating temperature
Storage temperature
Fuse:
Mains
Serial communication
Pump cable
In compliance with norms
Max altitude
Protection category
Internal use only
Weight (both models)
Installation category
Pollution degree
All connecting cables for remote I/O and serial connections, if longer than 3 m,
NOTE
must be of the shielded type.
TwisTorr Medium-TMP Rack Controller User Manual / 87-901-049-01 (B)
100 ÷ 240 Vac (±10 %), 1-phase
50 to 60 Hz
800 VA maximum
430 W (+/- 5%) maximum
200 W (+/- 5%) maximum
200 W (+/- 5%) maximum
5 °C to +40 °C
-20 °C to +70 °C
2 x T10 AH (slow blow) 250 V 5x20 fuses
RS232 cable with a 9-pin D type male connector and a 9-
pin D type female connector, and related software
(optional)
5 meters long fixed cable
EN 61010-1 (2010)
EN 61326-1 (2013)
Emission level : Class B for Residential Application
Immunity level : for Industrial Application
2000 m
IP 20
6.3 kg (13.9 lbs)
II
2
Technical Information
Controller Specifications
17
211/284