3 INSTALLATION
3.2.8
Wiring of a wall-mounted version, 12-30 VDC, without relay
Install the transmitter as described in § 3.1.3. Unscrew the 4 screws of the cover and uns-
crew the cable glands. Pass the cables through the cable glands by preferably respecting
the following allocation in order to ease the wiring:
Redox and Pt cables
Current output cable
Fig. 3.14 Allocation of the cable glands, wall-mounted versions
Connect the terminals according to the pin assignment below.
Connection of transmitter 8206 in a wall-mounted version without relay to a PLC is similar
to the connection of a panel-mounted version without relay (see Fig. 3.11).
Not connected
Temperature sensor
connection:
1 : red wire
2 : blue wire
3 : white wire
4 : green/yellow wire
PE : shielding of the cable
1
2
Pt1000
3
4
PE
Equipotential of the medium
Fig. 3.15 Wiring of the transmitter, wall-mounted version, 12-30 VDC, without relay
REDOX TRANSMITTER 8206
L+ (12-30 VDC)
1 2 3 4
NC
L+
L-
PE
Without
With
Iout
L+
L-
PE
Supply
12..30Vdc
1 2 3 4 PE
3A/230VAC
PT SENSOR
Fit here the cables using the supplied cable
clips (see § 3.2.3)
E-19-
Relay cables
12-30 VDC or 115/230 VAC
power supply cables
Connect the shielding of the
power supply cable
CURRENT
SOURCE
SINK
PE
PH SENSOR
redox sensor connec-
tion