Test voltages:
Accuracy:
Test current:
Number of measurements
according to EN 615577-2: about 6000 (250 V)
The insulation resistance measuring function is disabled when the input voltage
is 30 V AC/ DC or greater.
7.2­­ Low­­resistance­­measuring­­range­­(switch setting Ω 200 mA)
Measuring range
0 Ω
Analog bargraph display:
Test current:
Open-circuit voltage:
Number of measurements
according to EN 615577-4:
Zero compensation of the measuring cables up to 10 Ω.
Input protection:
The low resistance measuring function is disabled when the input voltage is
30 V AC/ DC or greater.
7.3­­ Resistance­­measuring­­range­­and­­acoustic­­continuity­­testing
(Switch setting:  )
Measuring range
000 Ω
Analog bargraph display:
Accuracy:
The internal buzzer sounds when the resistance R  30 .
Test current:
Input protection:
7.4­­­­ V oltage­­measuring­­range­­
(Switch setting: V
For AC (alternating voltage) the measured value is obtained by mean value
rectification and displayed as r.m.s. value. The specified accuracy for AC
measurements is based on the sinusoidal waveform.
Measuring range
000 V AC/ DC
Frequency range:
Analog bargraph display:
Accuracy:
Optical warning that a dangerous voltage is applied as from 30 V AC/ DC.
7.5­­ Measuring­­the­­battery­­capacity­­
(switch setting:
In accordance with EN 61557, the battery capacity is measured by connecting a
simulated load. Percentage indication from 0 - 100 % in 10 % steps.
8.­­­­ Making­­measurements­­with­­the­­BENNING­­IT­­100
8.1­­­­ P reparations­­for­­making­­measurements
Operate and store the BENNING IT 100 only at the specified storage and oper-
ating temperatures. Avoid continuous insolation.
-
C heck the nominal voltages and nominal current on the safety measuring cables.
The nominal voltage and current ratings of the safety measuring cables included
in the scope of delivery correspond to the ratings of the BENNING IT 100.
-
C heck the insulation of the safety measuring cables. Discard the safety
measuring cables immediately if the insulation is damaged.
-
C heck safety measuring lines for continuity. If the conductor in the safety
measuring line is interrupted, the safety measuring line must be quaran-
tined immediately.
-
B efore - at the rotary switch  - a different function is selected, the safety
measuring lines must be disconnected from the measuring point.
11/ 2006
250 V, 500 V, 1000 V
+ 20 %, - 0 %
1 mA
about 5000 (500 V)
about 3500 (1000 V)
Resolution
0,0 Ω
0 to 100 Ω and infinity
> 200 mA, 0 - 2 Ω
6 V DC
about 4000
0.5 A (1000 V AC/ DC) - fuse, quick blow
Resolution
 Ω
0 to 10 k and infinity
10 %
1 mA
0.5 A (1000 V AC/ DC) - fuse, quick blow
)
Resolution
 V
up to 400 Hz
0 to 1000 V
10 %
)
BENNING IT 100
Accuracy
 % ±  digits
Accuracy
 % ±  digits
Accuracy
 % ± 5 digits
Overload protection
000 V AC/ DC
Overload protection
000 V AC/ DC
Overload protection
000 V AC/ DC
5