Stereo microscope
Stemi 508
3.3.1.4
Placing reversible eyepiece cup onto the eyepiece
The eyepieces feature rubber protection rings for spectacles to avoid the spectacles being scratched.
These rings may be replaced at choice by reversible eye cups (Fig. 11/1).
To this effect remove the protection rings for spectacles (Fig. 11/2) from the eyepieces and position
the eye cups (Fig. 11/1).
3.3.1.5
Insert the eyepiece plate into the
eyepiece
The adjustable eyepieces are intended for use with
eyepiece plates.
Unscrew stop (Fig. 11/6) from the eyepiece
(Fig. 11/3).
Pull locking ring (Fig. 11/5) out.
Insert eyepiece plate (Fig. 11/4).
Insert locking ring and re-fasten stop.
If eyepiece plates are inserted by the
customer, it must be ensured that the
writing
is
insertion into the eyepiece, and true to
side after insertion.
Fine cotton gloves should be worn to insert or replace eyepiece plates.
The slight image offset which is due to the additional glass distance is taken into account on the diopter
scale by the zero position not being indicated by the white dot (Fig. 11/W), but by a red dot (Fig. 11/R).
Eyepieces with pre-inserted eyepiece plates are available directly from ZEISS.
09/2017
INSTALLATION
Mounting optional components
mirror-inverted
before
435064-7144-008
Fig. 11
Inserting the eyepiece plate
ZEISS
23