Controller K LED
•
Connect controller K to the power supply using the power unit (Fig. 18/1).
•
Align swivel arm stand (Fig. 18/6) to enable an ergonomic working position.
•
Turn the knob (Fig. 18/5) on the controller K to switch the ring illuminator ON and adjust illumination
intensity by turning the knob.
•
Place specimen in the illuminated specimen field.
•
Set the stereo microscope to its initial setting, see Section 4.1 on page 25.
28
OPERATION
Setting reflected-light illumination
4.2.4
The Controller K (Fig. 18/4) is used to activate the
ring illuminator in case of applications with boom
stands B or U or additional stands.
•
Mount the stands B and U including the
fastening of the Stemi 508 body to the Stemi
mount
instructions for the boom stands.
•
Screw-fasten the front lens system (Fig. 18/7) to
the microscope body.
•
Clamp ring illuminator to the front lens system.
•
Secure the adhesive side of the magnetic plate
(Fig. 18/7) to the swivel arm stand.
•
Plug the connection cable supplied with the
device (1 m long) (Fig. 18/2) into the socket
(Fig. 18/8) on the ring illuminator and into the
socket
controller K.
•
Place controller K (Fig. 18/4) next to the
magnetic plate (Fig. 18/7).
435064-7144-008
Stereo microscope
Controller K LED
as
described
in
Before using the front lens systems 0.3x
or 0.4x, fasten the retainer ring A53
mounting the ring illuminator to the
front lens system.
(Fig. 18/3)
on
the
Stemi 508
the
operating
back