PROCEDIMIENTO REDOX (mV)
Put in short circuit the BNC connector. Using a
metal wire, connect the core pin with the external
cylinder part.

AL A RM S set t in g

Overdosing alarm: For any kind of problems may occur in the plan (wrong calibration, dirty or broken probe, etc.) if the
pump doesn't reach the Setpoint in OVER ALARM TIME, the pump stops the dosing. On the display Alarm message will flash,
Alarm Led will flash.
To enter in ALARM menu follow the procedure
described in paragraph - PUMP SETUP (pag.81).
When ALARM menu is reached, press OK to enter
in submenus

OV E R DO SI N G AL A RM set t in g

On the display will appear OVER and time counter
(h:min) will flash. Press + and - to adjust the time,
over which the pump will go in alarm.
Press OK to confirm and to save data.
This will end the alarm setting.
Press ESC to return in measure mode.
ENGLISH
84
POINT1: Adjust the value on the display
up to reach 0 using + and – buttons
Press OK to save the data
On the display will appear POINT 2
Dip the electrode in a buffer solution. We
suggest 250, 475 or 650mV
Adjust the value on the display up to
reach the buffer solution value using +
and - buttons.
Press OK to save the data.