Parameter
Tolerance d
Clean
Printout
6.3.1.2 Perform the test
Start the test
§
The density meter is set up to work with the sample pump.
§
The measuring cell has been cleaned.
§
The piston is in the lowermost position.
1 Press the Menu key.
2 Navigate to Test.
ð The Test window opens.
3 Press the Start key.
ð The home screen opens.
4 Press the Start key.
ð The Standard ID window opens.
5 Enter the identification.
6 Press the OK key.
Rinse the measuring cell
1 Immerse the filling tube in the water.
2 Press the OK key.
ð The sample pump starts and stops when the piston is in the uppermost position.
3 Hold the filling tube over the waste beaker.
4 Press the OK key.
ð The sample pump starts and stops when the piston is in the lowermost position.
5 Clean the end of the filling tube with a clean tissue.
6 Repeat the steps once.
Fill the measuring cell and measure the density
1 Immerse the filling tube in the water.
2 Press the OK key.
ð The sample pump starts and stops when the piston is in the uppermost position.
3 Make sure there are no bubbles in the measuring cell.
ð The measurement starts.
ð When the background of the measuring ribbon changes color, the measurement is finished.
ð The measured value is saved in the results.
4 Press the OK key.
Drain the measuring cell
1 Hold the filling tube over the waste beaker.
2 Press the OK key.
ð The sample pump starts and stops when the piston is in the lowermost position.
3 Clean the end of the filling tube with a clean tissue.
Handheld Density Meter
Setting
3
0.0002 g/cm
Drain
None
Explanation
The result is marked as passed if it falls within the
following range.
d nominal ± Tolerance d
Users are prompted to drain the measuring cell.
The result is not printed.
17