The firmware update requires to use the « flash magic » software that can be download
(free) from this adress : www.flashmagictools.com
After installation, please configure the software on the following way :
Process
1 – Connect the D70A with the cable cable ref 804.2201 to the computer
2 – Start the D70A
3 – Configure the flash magic sotware according to the above picture.
4 – Click on "Start" It takes from 2 to 5 min depending on the Baud rate selected. During the
update, the screen becomes blurred.
5 - The D70A restart automatically when it is finished.
6 - Reset the D70A by following the procedure on the previous chapter.
The LPC4357 can be found on
the "ARM Cortex" category
after clicking on
« Select ».
If you are connected directly
on a serial port, you can use
the baud rate 115200bauds.
If you use a USB/RS232 con-
vertor, use 38400 bauds.