sylvac D70A Manuel D'utilisation page 56

Unité d'affichage pour mesure à air
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

6.2. Display
After pressing the icon «
This window allows to define if 1 or 2 measure are displayed on the screen and on which format.
(Needle = galva, horiziontal bargraph or value only)
The +/- keys appears on the needle indicator (galva) and allow to move the needle like with the screw on an old analogue
display.
6.3. Setup
After pressing the icon «
From this windows you can calibrate each channel. (the D70A must be calibrated with 2 maters).
Before mastering, this
field indicates the pressure
in BAR. After it should in-
dicate 0.0000mm when the
min master is in measuring
position, and the dimension-
al difference between the
2 masters when the MAX
master is inserted.
MIN and MAX master input
area.
display
», the window below appears.
The auto-switch function allows to call the corresponding program
by a simple probe movement. Example:
- Move probe 1, part 1 is displayed.
- Move probe 2, part 2 is displayed.
setup
», the windows below appears.
Channel 1 and
channel 2 (only
for D70A)
56
Left, simple
Left, double :
Galva :
None, simple :
None, double :
Allows to apply a
coefficient to the
channel. The
Defaut value is 1.
Press the button when the
master value has been writ-
ten and that the master is in
leasuring. This button be-
comes green for 1s after
validation.