7.6 Rechargeable battery operation (optional)

ATTENTION

7.6.1 Recharge battery

The rechargeable battery package is charged via the delivered power cable.
Before the first use, the rechargeable battery package should be charged by connecting
it to the mains power cable for at least 15 hours.
To save the rechargeable battery, in menu (see chap.0.) the automatic switch-off
function < Autoff> can be activated.
If the capacity of the rechargeable battery is exhausted, in the display <instab> will
appear. Connect the power cable as soon as possible to load the rechargeable battery.
Charging time until complete recharging is approx. 10 h.
16
The rechargeable battery and the battery match with each
other. Only use the delivered mains adapter.
Do not use the balance during the loading process.
The rechargeable can only be replaced by the same or by a
type recommended by the manufacturer.
The rechargeable battery is not protected against all environmental
influences. If the rechargeable battery is exposed to certain
environmental influences, it may set on fire or explode. Persons may be
injured or material damage may occur.
Protect the rechargeable battery against fire and heat.
Do not bring the rechargeable battery in contact with fluids, chemical
substances or salt.
Do not expose the rechargeable battery to high pressure or
microwaves.
Under no circumstances the rechargeable batteries and the charging
unit may be modified or manipulated.
Do not use a defective, damaged or deformed rechargeable battery.
Do not connect or short-circuit the electrical contacts of the
rechargeable battery with metallic objects.
Liquid may squirt out from a damaged rechargeable battery.
If the liquid gets into contact with the skin or the eyes, the skin and the
eyes may be irritated.
Ensure the correct polarity when inserting or changing the rechargeable
battery (see instructions in the battery compartment)
The rechargeable battery operation is overridden when the mains
adapter is connected. For weighing in mains operation > 48 hrs., the
rechargeable batteries must be removed! (Danger of overheating).
If the rechargeable battery starts to smell, being hot, changing
the colour or being deformed, it must be immediately unplugged
from mains supply and from the balance if possible.