Audio signal:
The acoustic signal depends on the menu setting
< setup  beeper >, see chap. 13.3.1.
Procedure:
1. Define upper and lower limit value:
Make sure that the balance is in Checkweighing mode. If necessary, switch over
with the
Select weighing unit for checkweighing mode
Press the TARE button and keep it pressed until the first menu item < limit >
will be displayed.
Confirm with  button. < limlow > will be displayed.
Press  button to confirm, the numeric input window for entering the lower limit
value will appear. Enter the lower limit value for the target piece quantity (numeric
input see chap. 3.2.2) and confirm with  button.
The balance returns to the < limlow > menu.
Use the navigation keys  to select the setting < limupp > and confirm with 
button.
The numeric input window for entering the upper limit value < will appear. Enter the
upper limit value for the target piece quantity (numeric input see chap. 3.2.2) and
confirm with  button. < limupp > will appear.
Press repeatedly  button to exit menu.
Finished the setting works, the weighing balance is ready now for check weighing.
2. Start tolerance check:
Place the weighed material and check by means of the tolerance marks / acoustic
signal if the weighed material is within the defined tolerance.
Load below specified
tolerance
FKB-BA-e-2111
Load within specified
tolerance
Load exceeds specified
tolerance