Technical Information
Controller Connection
Controller Connection
P2 – Fan
The 24 Vdc, 100 mA maximum output voltage is present after Switch
ON of the controller. This is a dedicated output for the optional
Turbo-V 81-AG pump cooling fan (see orderable parts table). If a fan
is already installed, use the available adapter cable to connect to the
pump.
Figure 9
P1 – Vent
Figure 10
This is a dedicated 24 Vdc connector for control of the optional vent
valve (see orderable parts table). The vent valve can be driven by the
controller automatically or via a serial line. If a 110 Vac-powered
vent valve has already been installed, this valve must be replaced
with a new model (see orderable parts table).
156/192
P2 Fan Connector
P1 Vent Connector
Turbo-V 81-AG Navigator Controller User Manual / 87-901-004-01 (G)