Cleaning And Disinfection; Cleaning; Uv Disinfection; Troubleshooting - BRAND HandyStep S Mode D'emploi

Distributeur à répétition
Masquer les pouces Voir aussi pour HandyStep S:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 53

7 Cleaning and disinfection

7.1 Cleaning

The instrument is not autoclavable
The instrument is factory-calibrated and maintenance-free.
When the instrument is dirty, clean the outer surface with a moist cloth.
Use water or a mild detergent solution. Avoid acidic or aggressive clean-
ers.
Disassembly of the instrument is not permitted.

7.2 UV disinfection

The device is resistant to normal exposure to a UV disinfection lamp. The
device may change color due to the UV exposure.
UV disinfection recommendations:
Light spectrum
Wave length
Exposure time per cm
inactivation rate

8 Troubleshooting

Problem
PD-Tip piston cannot be
locked after the PD-Tip has
been inserted.
Undefined dispensing vol-
ume
9974 63
UV-C
220 nm ... 270 nm
2
and intended
2 s ... 300 s
Possible cause
Filling/locking lever is not
pushed down to bottom position
and is not swung out.
The stroke setting dial is not
properly engaged.
NOTICE
Corrective action
First press the ejector button and
remove the PD-Tip, then push the
filling/locking lever all the way
down and swing it out com-
pletely.
Ensure that the stroke setting dial
is securely locked into place.
7 Cleaning and disinfection
Operating manual
45

Publicité

Table des Matières
loading

Table des Matières