10 Technical data
10.1 Accuracy table
PD-Tips II from BRAND, 20 °C 'Ex'
PD-Tips
Volume range
sizes
(µl)
0.1 ml
2 - 10
0.5 ml
10 - 50
1.0 ml
20 - 100
1.25 ml
25 - 125
2.5 ml
50 - 250
5.0 ml
100 - 500
10.0 ml
200 - 1000
12.5 ml
250 - 1250
25.0 ml
500 - 2500
50.0 ml
1000 - 5000
A* = Accuracy, CV = Coefficient of Variation
Error limits in relation to the set dispensing volume depending on the PD-
Tip size, at equal temperature (20 ºC) of instrument, tip, ambient environ-
ment and distilled water, and with uniform, jolt-free handling. The test is
carried out in accordance with DIN EN ISO 8655-5. The nominal volume is
the maximum volume printed on the PD-Tip.
10.2 Dispenser tips from other manufacturers
We recommend using BRAND PD-Tips II; however, the HandyStep® S can
also be used with compatible dispenser tips from other manufacturers.
When using compatible dispenser tips, the user must verify their suitabil-
ity and carry out a function/volume test!
9974 63
A* ≤ ± %
stroke setting % of nominal
volume
1 2%
3 6%
5 10%
4.0
2.4
1.6
2.5
1.5
1.0
2.5
1.5
1.0
2.5
1.4
0.9
1.8
1.1
0.7
1.8
1.1
0.7
1.8
1.1
0.7
1.8
1.1
0.8
1.5
0.9
0.6
1.5
0.8
0.5
10 Technical data
CV* ≤ %
stroke setting % of nominal
volume
1 2%
3 6%
5 10%
6.0
3.0
2.0
2.5
1.5
1.0
2.0
1.2
0.8
2.0
1.1
0.7
1.5
0.9
0.6
1.5
0.9
0.7
2.0
1.2
0.8
3.2
2.0
1.4
3.0
1.5
1.0
5.0
1.8
1.2
Operating manual