Bair Hugger Model 505 Temperature Management Unit Operator's Manual
Specifications
Model 505
Physical Characteristics
DIMENSIONS
WEIGHT
MOUNTING
RELATIVE NOISE LEVEL
HOSE
FILTRATION SYSTEM
RECOMMENDED FILTER CHANGE
Temperature Characteristics
TEMPERATURE CONTROL
HEAT GENERATED
SYSTEM TIME TO
OPERATING TEMPERATURES
Safety System
THERMOSTAT
OVERCURRENT PROTECTION
ALARM SYSTEM
CERTIFICATIONS
CLASSIFICATION
100°F (37.7°C)
13 in. high x 10 in. deep x 11 in. wide
33 cm high x 25 cm deep x 28 cm wide
13.6 lb; 6.2 kg
IV pole clamp, bedrail hook with safety strap;
can be placed on hard surface
53 decibels
Detachable, flexible, washable; compatible
with the 241 fluid warming system
0.2µM level
Every 500 hours of use
Electronically controlled using a thermocouple
sensor
1800 BTUs (average)
~17 secs
Air temperatures reaching the patient are
approximately 2°C lower than the listed
temperatures.
Average temperatures at the end of the hose:
HIGH:
43° ± 3°C
MED:
38° ± 3°C
LOW:
32° ± 3°C
Independent bulb and capillary
Dual fused input lines
Over-heat: flashing red light with audible
alarm; heater shuts down
UL 544, CSA C22.2 No. 125, IEC 601-1,
IEC 601-1-2, EN55014, AS 3200.1990
Classified under IEC 601-1 Guidelines as Class I,
Type BF, Ordinary equipment, Continuous operation
109.4° ± 5.4°F
100.4° ± 5.4°F
89.6° ± 5.4°F