6. Commissioning, operation
To exit the submenu, press the [MENU] button.
Operation
On leaving the menu, the controller will revert to standby status. All information about the
EN
AUTO mode will be shown on the display.
VENT
6
Cycl.
6
/ 10
5
Start
1
2
4
5
Next cycle
End of cycle
The figure shows the flow chart for the AUTO mode. Based on the table, one will know
which cycles are running and the meaning of each.
Phase
Meaning
1
Wait for the start command with [SELECT]
2
Delay time can be defined in the configuration
Duration of the system zero point correction
4
7
Hold time at the zero point
9
Time to control up to the next step
10
Hold time
Delay time at the maximum target pressure
13
24
1
+10.000
mbar
Phase
1
Time
Press
-0.007
+
0%
+0.000
4
7
9
10
9
10
1
2
2
3
1
3
5 s
4
5
6
6
9
10
9
10 13 15 16 15 16 15 16 15 16 18
WIKA operating instructions, model CPC2000
Full scale
Pressure unit
Phase 1: it has been vented
(Pressure = 0)
Percentage display of the
current target pressure
Battery status
First of ten cycles
Operating mode