WIKA CPC2000 Mode D'emploi page 23

Masquer les pouces Voir aussi pour CPC2000:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

6. Commissioning, operation
3. Confirm with [SELECT].
For the AUTO mode, the submenu points RANGE, UNIT and STEPS are not
relevant.
4. With the [▲] or [▼] buttons, select the menu item SETTINGS.
5. Confirm with [SELECT].
The submenu item ZERO is not relevant.
6. In the submenu, select the item AUTO-MODE with the [▲] or [▼] buttons.
7. Confirm with [SELECT].
The following items can be set in the operating mode AUTOMATIC-MODE:
Menu item
RANGE
UNIT
STEPS UP
STEPS DOWN
CONFIGURATION 1
In the submenu item CONFIGURATION 1, further settings can be defined for the AUTO
mode.
Settings in the submenu item AUTO-MODE CONFIG 1
Menu item
CYCLES
T START
T STOP
T HOLD
T PAUSE
AUTO ZERO
WIKA operating instructions, model CPC2000
Meaning
Pressure range to be measured
Selection of the pressure unit (Pa, kPa, hPa, bar, mbar, psi, inH
inHg, mmHg, Torr)
The number of steps which will be taken to increase the pressure.
These step sizes are calculated by the controller itself.
The number of steps which will be taken to decrease the pressure.
These step sizes are calculated by the controller itself.
Further required settings for the AUTO mode can be set
AUTO-MODE CONFIG 1
-------------------------
CYCLES
T START
T STOP
T HOLD
T PAUSE
AUTO ZERO
Meaning
Here the number of cycles to be carried out in AUTO mode should be set.
Delay time to vent the sensor
Delay time after reaching the maximum defined pressure
Hold time until a new pressure value is reached
Delay time between two cycles
Automatic zero point correction at the completion of each cycle
001
00005 s
00005 s
00005 s
00005 s
ON
EN
O,
2
23

Publicité

Table des Matières
loading

Table des Matières