6. Commissioning, operation
6.9
SETUP settings (SETUP)
The SETUP menu items are made up of six submenu items.
EN
Menu item
RANGE
UNIT
STEPS
MODE
LANGUAGE
SETTINGS
Navigation within the menu:
[SELECT]: Acts to confirm the relevant setting
SELECT
[◀] or [▶] arrows: Setting of the active decimal place
[▲] or [▼] arrows: Changes between the different submenu items
26
MENU
-------------------------
RANGE.....:
UNIT......:
STEPS.....:
MODE......:
LANGUAGE..:
SETTINGS
Meaning
Setting of the pressure range (max. 100 % of the measuring span)
Selection of the pressure unit
Selectable pressure units:
Pa, kPa, hPa, bar, mbar, psi, inH
The selected pressure unit will be shown on the display.
Dependent upon the measuring range of the instrument, various units are
not available since they could not be displayed.
Setting of the step size in x %.
The step size for the STEP function, in x % of the set range, can be
changed.
Possible settings are from 1 ... 50 %.
Selection and setting of the operating mode
The following modes can be selected:
Measure "MEAS": Measurement of differential and gauge pressure in
■
the set pressure unit
Control "CTRL": Regulates to the specified set point automatically
■
Automatic "AUTO": A stored test sequence is processed
■
Selection of the display language
The language in which the menu will be displayed can be set here. There
is a choice between German, English, Spanish and French.
Here additional instrument settings can be made
+10.000
mbar
50%
MEAS
EN
-->
O, inHg, mmHg, Torr
2
WIKA operating instructions, model CPC2000