3. Safety
3.2
Intended use
The model CPC2000 low-pressure controller is used for testing and calibrating pressure
sensors.
EN
This instrument is not permitted to be used in hazardous areas!
The instrument has been designed and built solely for the intended use described here,
and may only be used accordingly.
The technical specifications contained in these operating instructions must be observed.
Improper handling or operation of the instrument outside of its technical specifications
requires the instrument to be taken out of service immediately and inspected by an
authorised WIKA service engineer.
Handle electronic precision measuring instruments with the required care (protect from
humidity, impacts, strong magnetic fields, static electricity and extreme temperatures,
do not insert any objects into the instrument or its openings). Plugs and sockets must be
protected from contamination.
The manufacturer shall not be liable for claims of any type based on operation contrary to
the intended use.
3.3
Improper use
DANGER!
Danger to life from explosion!
Improper use of the low-pressure controller may result in a risk of explosion,
which may lead to serious injury or death, as the built-in lithium-ion
rechargeable battery may explode.
▶
Do not burn or heat up the low-pressure controller.
▶
Never store the low-pressure controller close to fire, ovens or other
high-temperature locations.
▶
Never place the low-pressure controller in a hermetically sealed container.
In some cases hydrogen or oxygen may be discharged from the lithium-ion
rechargeable batteries, which may result in rupture, fire or explosion.
WARNING!
Injuries through improper use
Improper use of the instrument can lead to hazardous situations and injuries.
▶
Refrain from unauthorised modifications to the instrument.
▶
Do not use the instrument within hazardous areas.
▶
There must be no external pressure acting on the CPC2000.
▶
Observe the operating parameters in accordance with chapter
11 "Specifications".
8
WIKA operating instructions, model CPC2000