Les langues disponibles

6. Commissioning, operation
Operation
On leaving the menu, the zero point of the internal sensor is adjusted, dependent upon
the setting. The instrument then begins to measure the subsequent pressure.
EN
MEAS
6
+00.002
5
Using the [ZERO] button in MEAS mode, the instrument's zero point can be
compensated at any time.
If the connected pressure should exceed the permissible maximum load
(125 %), the internal sensor will be closed off using a solenoid valve. The
display will then show ERROR. Using the [SELECT] button, the instrument
can be opened and unlocked again.
6.7

Operating mode AUTOMATIC-MODE "AUTO"

In AUTO mode, a pressure profile can be stored. This is especially useful when multiple
pressure sensors or pressure switches are to be calibrated with the same values. The
mode also enables a defined profile to be carried out several times in succession.
Setting operating mode AUTOMATIC-MODE
1. Press the [MENU] button.
2. Confirm the menu item MODE with [SELECT] and press the [▲] / [▼] buttons until
AUTO is displayed.
22
1
+10.000
mbar
+
0%
+0.002
4
AUTOMATIC-MODE
-------------------------
RANGE
UNIT
STEPS UP
STEPS DOWN
CONFIGURATION 1
Full scale
1
Pressure unit
2
2
Measured value
3
Set point in % of full scale
4
3
Battery status
5
Operating mode
6
0...+10.000
mbar
005
005
-->
WIKA operating instructions, model CPC2000