6. Commissioning, operation
Operation
On leaving the menu, the zero point of the internal sensor is adjusted, dependent upon
the setting. The instrument then begins to measure the subsequent pressure.
EN
MEAS
6
+00.002
5
Using the [ZERO] button in MEAS mode, the instrument's zero point can be
compensated at any time.
If the connected pressure should exceed the permissible maximum load
(125 %), the internal sensor will be closed off using a solenoid valve. The
display will then show ERROR. Using the [SELECT] button, the instrument
can be opened and unlocked again.
6.7
Operating mode AUTOMATIC-MODE "AUTO"
In AUTO mode, a pressure profile can be stored. This is especially useful when multiple
pressure sensors or pressure switches are to be calibrated with the same values. The
mode also enables a defined profile to be carried out several times in succession.
Setting operating mode AUTOMATIC-MODE
1. Press the [MENU] button.
2. Confirm the menu item MODE with [SELECT] and press the [▲] / [▼] buttons until
AUTO is displayed.
22
1
+10.000
mbar
+
0%
+0.002
4
AUTOMATIC-MODE
-------------------------
RANGE
UNIT
STEPS UP
STEPS DOWN
CONFIGURATION 1
Full scale
1
Pressure unit
2
2
Measured value
3
Set point in % of full scale
4
3
Battery status
5
Operating mode
6
0...+10.000
mbar
005
005
-->
WIKA operating instructions, model CPC2000