batt
appears in the display?
– The rechargeable battery is so severely discharged that weighing is no longer possible.
Recharge the battery.
...
St0p
appears in the display?
– Maximum load has been exceeded.
... the display flashes?
– If you have just activated the Hold or Tare function, please wait until the scale has de-
tected a steady weight value. The display will then stop flashing.
– If there has been no activation of a function beforehand, remove the load from the scale
0. 0
and wait until
... the display shows
– The ambient temperature of the scale is too high or too low. Place the scale in an
ambient temperature between +10 °C and +40 °C. Wait about 15 minutes for the
scale to adapt to the ambient temperature and then weigh again.
E
... the display
and a number appear?
– Switch off the scale with the Start key and start the scale again. The scale will then
work normally again.
If this is not the case, interrupt the power supply by briefly removing the rechargeable
battery. If this measure is equally unsuccessful, inform the Service department.
9. Maintenance
Your seca scale leaves the factory with an accuracy greater than ± 0,15 %. To ensure that
this accuracy is maintained, the product must be carefully set up and regularly maintained.
Depending on the frequency of use, we recommend servicing every 3 to 5 years. Please
ensure that a qualified service agent is used. If in doubt, contact seca Customer Service for
further information.
10. Technical data
Dimensions: (width, length, height)
Size of platform
Height of platform
Weight
Temperature range
EN 60 601-1:
insulated equipment, Protection Class II
medical electrical equipment type B
Accuracy
Colour
Maximum load
32
is displayed, then weigh again.
?
750 x 750 x 1140 mm / 30 x 30 x 45 ins
600 x 600 mm / 23.6 x 23.6 inch
45 mm / 1.77 inch
approx. 24 kg / 53 lbs
+10 °C to +40 °C
greater than ± 0,15 %
black/white
300 kg / 660 lbs / 47 sts