0% UT; 0,5
cycle
At 0°, 45°, 90°,
135°, 180°, 225°,
Voltage dips,
270° and 315°
short
interruptions and
0% UT; 1 cycle
voltage
and
variations on
70% UT; 25/30
power supply
cycles
input lines
Single phase: at
IEC 61000-4-11
0°
0% UT; 250/300
cycle
NOTE: U
is the AC mains voltage prior to application of the test level.
T
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the product is used exceeds
the applicable RF compliance level above, the product should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 1 V/m.
0% UT; 0,5 cycle
At 0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°
0% UT; 1 cycle
and
70% UT; 25/30 cycles
Single phase: at 0°
0% UT; 250/300 cycle
57