3V in 0,15 MHz
to 80 MHz
Conducted
6V in ISM and
disturbances
amateur radio
inducted by RF
bands between
fields
0,15 MHz and 80
MHz
EN 61000-4-6
80% AM at 1
kHz
Professional
Healthcare
Radiated RF
environment
electromagnetic
3 V/m
field
80 MHz to 2,7
EN 61000-4-3
GHz
80% AM at 1
kHz
385 MHz - 27 V/
m
450 MHz - 28 V/
m
Proximity fields
710, 745, 780
from RF wireless
MHz - 9V/m
communications
810, 870, 930
equipment
MHz - 28 V/m
EN 61000-4-3
1720, 1845,
1970, 2450 MHz
– 28 V/m
5240,5500, 5785
MHz - 9V/m
±1kV SIP/SOP
ports
Electrical fast
±2kV AC port
transient/burst
100kHz
EN 61000-4-4
repetition
frequency
Power frequency
30A/m
Magnetic field
50 Hz or 60 Hz
EN 61000-4-8
±0,5kV ±1kV;
±2 kV, AC
Mains, Line to
Surge
Ground
IEC 61000-4-5
±0,5kV ±1kV,
AC Mains, Line
to Line
56
3V in 0,15 MHz to 80 MHz
6V in ISM and amateur radio
bands between 0,15 MHz and
80 MHz
80% AM at 1 kHz
Professional Healthcare
environment
3 V/m
80 MHz to 2,7 GHz
80% AM at 1 kHz
385 MHz - 27 V/m
450 MHz - 28 V/m
710, 745, 780 MHz - 9V/m
810, 870, 930 MHz - 28 V/m
1720, 1845, 1970, 2450 MHz
– 28 V/m
5240,5500, 5785 MHz - 9V/m
±1kV SIP/SOP ports
±2kV AC port
100kHz repetition frequency
30A/m
50 Hz
±0,5kV ±1kV; ±2 kV, AC
Mains, Line to Ground
±0,5kV ±1kV, AC Mains,
Line to Line
Portable and mobile RF
communications equipment
should be used no closer to
any part of the product,
including cables, than 1.0m,
if the transmitter's output
a
power rating exceeds 1W
Field strengths from fixed RF
transmitters, as determined
by an electromagnetic site
survey, should be less than
the compliance level in each
b
frequency range
.
Interference may occur in the
vicinity of equipment marked
with this symbol:
Mains power supply should
be that of a typical
commercial or hospital
environment.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
.