Setting the Calibration Gas Value

Before you can proceed to the calibration, you must make sure that the
calibration gas concentration value in the IAQ Probe is the same as
what is indicated on the calibration gas cylinder.
1
Press Mode/Esc to access the main menu.
2
Use the arrows to scroll to Gas and press Enter to confirm.
3
Use the arrows to scroll between CO
select. The selected sensor concentration is displayed.
4
Press Enter. The concentration is editable when flashing.
5
Use the arrows to change the value (increase or decrease
according to the information on the cylinder) Enter to confirm.
6
Press Mode/ESC to return to the main menu.
7
Press Mode/ESC to exit the menu.