Fischer Cone Biopsy Excisor
•
Loop excision procedures performed with small diameter wire loop
electrodes produce multiple small pieces of cervical tissue and
provide a less acceptable tissue specimen for histopathologic
analysis. The effectiveness of the procedure and influence of
electrodesign are not completely understood.
•
Larger lesions involving multiple quadrants of the cervix are more
difficult to remove with either the small or large diameter loop
electrodes.
SMALL DIAMETER LOOP VS LARGE DIAMETER LOOP
The histological quality of the specimens obtained using the small
diameter loop was inferior to that obtained with the larger diameter loop.
This is because more of the epithelium comes in direct contact with the
small diameter electrode than with the large diameter electrode and that
part of the electrode is not insulated. This results in more thermal damage
to the tissue. In 7% of the samples, the amount of thermal damage
precluded histological evaluation of the excised tissue. In addition, since
multiple, often irregular, strips of epithelium are excised from the cervix, it
is often difficult for the pathologist to orient the specimens for optimal
histopathological examination.
THERMAL INJURY AND DEFECTS OF TISSUE TREATED WITH LOOP
ELECTRODES
Possible injury to cervical tissue may include: (I) thermal coagulation
injury of the cervix, up to one-third the thickness of normal epithelium of
the cervix, (II) fragmentation of squamous epithelium of the cervix
attributable to long exposure periods along the excision site that allows
heat to dissipate laterally, and (III) partial coagulation of the endocervical
epithelium because of lateral radiation of heat.
Loop electrosurgical excision procedures may also produce thermal
defects at the periphery of the excised tissue and may make
histopathological interpretation difficult or impossible and may prevent
accurate diagnosis and the evaluation of the need for further treatment.
RELATED PRODUCTS
•
KH1000
•
6050P
•
6040
•
6084
•
6083
®
LEEP System 1000 Workstation
Return Pad - single-use
Handpiece - single-use
Tubing set - single-use
Reducer - single-use
• Instructions for Use • English (
•
52560
•
6082
•
6060
7
)
continued
Prefilter - single-use
ULPA Filter
LEEP RediKit
®