Operation; Selecting The Curing Time - Ivoclar Vivadent BluephaseN M Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 15

4. Operation

Disinfect contaminated surfaces of the curing light as
well as the light probe and anti-glare cone before each
use. Furthermore, make sure that the light output emit-
ted by the curing light permits adequate polymerization.
For that purpose, inspect the light probe for contami-
nation and damage and check the light intensity at
regular intervals (e.g. with the Ivoclar Vivadent radio-
meter "Bluephase Meter").

Selecting the curing time

The curing time can be
individually set using the time
selection button. You may
choose between 10, 15 and
20 seconds.
Curing times
Restorative materials
Indirect restorations /
cementation materials
Adhesives
Temporary materials
Miscellaneous
1) Applies to a maximum layer thickness of 2 mm and provided that the Instructions for Use of
the respective material do not state any other recommendation (might be the case e.g. with
dentin shades)
2) Applies to a maximum layer thickness of 4 mm and provided that the Instructions for Use of
the respective material do not state any other recommendation (might be the case e.g. with
dentin shades)
Composites
• 2 mm
1)
IPS Empress Direct / IPS Empress Direct Flow
Tetric EvoCeram / Tetric EvoFlow
Tetric / Tetric Basic White / Tetric N-Ceram / Tetric N-Flow
• 4 mm
2)
Tetric EvoCeram Bulk Fill / Tetric N-Ceram Bulk Fill
Tetric Basic White
Heliomolar / Heliomolar HB / Heliomolar Flow
All conventional composites
Compomers
3)
Compoglass F / Compoglass Flow
Variolink II Base
/ Variolink Veneer / Variolink N Base
4)
Variolink N Clear Veneer / Variolink Ultra
Dual Cement
/ Variolink II
5)
5)
AdheSE / ExciTE F / Tetric N-Bond / Tetric N-Bond Self-Etch /
Te-Econom Bond
Telio CS Link / Systemp.link
Systemp.inlay/onlay / Telio CS Inlay/Onlay / Fermit / Fermit N
Telio Stains
Telio Add-On Flow
Helioseal / Helioseal F / Helioseal Clear
Heliosit Orthodontic
Vivaglass Liner
IPS Empress Direct Color / IPS Empress Opaque
MultiCore Flow / Multicore HB
Observe the Instructions for Use of the material applied
when selecting the curing time. The curing recommenda-
tions for composite materials apply to all shades and, if
not mentioned otherwise in the Instructions for Use of
the relevant material, to a maximum layer thickness of
2 mm. Generally, these recommendations apply to
situations where the emission window of the light probe
is placed directly over the material to be polymerized.
Increasing the distance between the light source and the
material will require the curing time to be extended
accordingly. For instance, if the distance to the material
is 9 mm, the effective light output is reduced by approx.
50%. In this case, the recommended curing time has to
be doubled.
/
4)
5)
3) Applies to a maximum layer thickness of 3 mm
4) Applies to light-curing (use of the base past only)
5) Applies to dual-curing
800 mW/cm
± 10%
2
15 s
15 s
20 s
20 s
40 s
per mm ceramic:
10 s per surface
10 s
15 s per surface
10 s
15 s
20 s
20 s
30 s
40 s
9

Publicité

Table des Matières
loading

Table des Matières