8. Servicing/recalibration
Have the scale recalibrated by authorised persons in accordance with national legal re-
quirements. The year of initial calibration can be found behind the CE symbol on the rating
plate above the number of the appointed office, 0109 (Hessen Weights & Measures Of-
fice). Recalibration is essential if one or more security marks are damaged or the calibra-
tion counter does not match the number on the applicable calibration counter mark (cf.
page 26).
We recommend having your scale serviced by your local service agent before having it
recalibrated. seca Customer Service will be pleased to give you further assistance.
9. Technical data
Maximum load
Minimum load
Graduations
seca 959:
0 up to 150 kg:
150 up to 200 kg:
seca 958:
0 up to 100 kg:
100 up to 200 kg:
Dimensions
height:
width:
depth:
Weight
Temperature range
Power supply
Power consumption
Calibration as per Directive
90/384/EEC
seca 959:
seca 958:
Medical product as per Directive
93/42/EEC
EN 60 601-1:
Insulated equipment, Protection Class II:
medical electrical equipment type B:
Model 958/959
200 kg
1 kg
50 g
100 g
100 g
200 g
920 mm
565 mm
810 mm
approx. 24.5 kg
+10 °C to +40 °C
rechargeable batteries/mains unit
approx. 25 mA
approved for medical use, Class III
approved for medical use, Class IIII
Class I
31