Hach MET ONE 6000P Manuel D'utilisation page 14

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 24
®
• PC with Windows
2000 Professional, Windows XP Professional,
Windows Vista (32-bit), Windows 7 (32-bit or 64-bit in XP emulation
mode)
• USB to RS232 adapter if the PC does not have a RS232 port
1. Make sure that Microsoft .Net Framework is installed on the PC. If
not installed, open the dotnetfx.exe file on the setup utility CD and
install the application.
Note: The user must be logged on to the PC as an Administrator.
2. Copy and paste the SetupUtility.exe file from the setup utility CD to
the PC.
3. Connect the Service port cable to the Service port on the instrument
and a COM port on the PC.
Configure the instrument
Use the setup utility software to configure the parameters kept on each
instrument. When power is applied to the instrument, it looks for a new
configuration. If a new configuration is not found, the previously saved
configuration is used.
1. Open the SetupUtility.exe file that is installed on the PC to start the
setup utility program.
2. Select the Basic Setup tab.
3. Find the Port field on the right side of the window. Select the COM
port on the PC to which the instrument is connected.
4. Click Read Instrument. The utility reads the data that is saved on
the instrument.
5. Make sure that the data in the Instrument Information section is
correct (model number, communication option, firmware version and
communication address, if applicable).
14 English
6. In the General section, select the settings.
Option
Description
Count Mode
Sets the count mode. Does not affect the analog
output of analog units.
Differential—The particle counts shown for each
channel are the counts for each channel size.
Cumulative (default)—The particle counts shown
for each channel are the counts for each channel
size plus the larger channel sizes. For example, if
the channel is 0.3 µm, particles that are 0.3 µm and
larger in size are included in the count.
Sample Timing:
Sets the length of time for each sample (default =
Sample
00:01:00 = 1 minute).
Sample Timing: Hold
Sets the length of time that data collection stops
after samples are taken (default = 00:00:00).
Count Cycles
Sets the number of samples taken before data
collection stops and the hold time starts (0 =
continuous sampling).
Slave
Do not change (default = 1).
Address/Location ID
Comm Timeout
Sets the number of seconds after a communication
failure before a communication (Comm) alarm
occurs. To disable communication alarms, set to 0.
For instruments with the analog output feature, set
to 0.
Location Name
Sets a unique identifier for the instrument.
System Date/Time
Sets the date (YYYY/MM/DD) and time
(HH:MM:SS, 24-hour format).
Moving Cumulative
Sets the number of sample counts for Channel 1 or
Counts
Channel 2 that are added together and shown in
Channel 3 and Channel 4. Channel 3 shows the
cumulative counts for Channel 1. Channel 4 shows
the cumulative counts for Channel 2.
Store Partial Records
Enables partial sample data to be saved to the
buffer. Partial sample data occurs when a sample
is stopped before it is completed.