INSTALLATION
Dental Lamp "MAIA", version S/TS
Check that the packaging contains the following components:
- Dental lamp (requested version)
- Envelope with lever + key
- Operating Manual
Lamp assembly, dental chair version
- Install the lamp by inserting the terminal lamp pin into the specifi c hole on the dental chair.
Long arm 855 mm
Long arm 550 mm
1) POWER CORD:
- LAMP WITH TRANSFORMER; it must be directly powered by the power-line
voltage corresponding to characteristics printed on the information plate or in the
technical specifi cations of the manual.
- Ensure that the main switch complies with IEC/EN 61058 standards.
- LAMP WITHOUT TRANSFORMER; it must be powered by low voltage alternate
current (17 - 24V AC) using a safety transformer that meets EN 60601-1 standards.
Fitting the ceiling – wall – fl oor light
- The applications are not supplied with the lamp.
40°
50°
- For fi tting the individual applications follow the relevant installation manual.
18
CONNECTIONS
AND
Clean the device before use
(see section Cleaning the device)
The device must be installed by specialised technicians.
Disconnect power supply during installation.
CEILING
750
L=H-1700-20
SAFE WORKING LOAD MINIMUM BREAKING LOAD
2.92 Kg
2.56 Kg
WALL
40°
50°
23.5 Kg.
20.5 Kg
FLOOR
750
40°
50°