The lamp MAIA is designed to function in an electromagnetic environment in which irradiating RF disturbances are under control.
The client or operator of the unit can contribute toward preventing electromagnetic interferences by ensuring a minimum distance
between mobile and portable RF communication devices (transmittors) and the dental unit, as recommended below, depending on
the maximum output power of the radiocommunication devices.
Maximum nominal
output power
of the transmitter
For transmittors whose maximum nominal power is not listed above, the recommended distance d in metres (m) can be calculated
by using the applicable equation for the transmittor frequency, with P as maximum nominal output of the transmittor in Watts (W),
depending on the manufacturer.
Notes:
The highest frequency interval is applied at 80 MHz and 800 Mhz.
These guidelines might not apply to all situations. Electromagnetic propagation is infl uenced by absorption and refl ection of struc-
tures, objects and persons.
16
Recommended distances between portable and mobile radiocommunication
150 kHz to 80 MHz
d = 1,2 √P
W
0,01
0,12
0,1
0,38
1
1,2
10
3,8
100
devices and the dental unit
Distance for transmittor frequencies (m)
80 MHz to 800 MHz
d = 1,2 √P
0,12
0,38
1,2
3,8
12
12
800 MHz to 2,5 GHz
d = 2,3 √P
0,23
0,73
2,3
7,3
23