Warning
Description
PH TOO LOW
The measured pH is < 1.
ORP TOO LOW
The measured ORP value is < –2100 mV.
OFFSET TOO HIGH The offset is > 8 (pH) or 200 mV (ORP).
OFFSET TOO LOW
The offset is < 6 (pH) or –200 mV (ORP).
SLOPE TOO HIGH
The slope is > 60 (pH)/1.3 (ORP).
SLOPE TOO LOW
The slope is < 54 (pH)/0.7 (ORP).
TEMP TOO HIGH
The measured temperature is >100 °C.
TEMP TOO LOW
The measured temperature is < 0 °C.
CAL OVERDUE
The number of days since the sensor was
calibrated is more than the calibration
interval.
REPLACE SENSOR The number of days the sensor has been
in operation is more than the sensor
replacement interval.
NOT CALIBRATED
The sensor has not been calibrated.
FLASH FAILURE
External flash memory failed.
A ELEC TOO HIGH
The active electrode impedance is >
800 MΩ.
A ELEC TOO LOW
The active electrode impedance is <
15 MΩ.
R ELEC TOO HIGH
The reference electrode impedance is >
800 MΩ.
R ELEC TOO LOW
The reference electrode impedance is <
15 MΩ.
CAL IN PROGRESS A calibration was started but not
completed.

6.5 Event list

The event list shows current activities such as configuration changes, alarms, warning conditions,
etc. To show the events, push menu and select DIAGNOSTICS > [Select Sensor] > EVENTS. A list
of possible events is shown in
downloaded from the controller. Refer to the controller documentation for data retrieval options.
Event
CAL READY
CAL OK
TIME EXPIRED
NO BUFFER
Table 4 Warning messages (continued)
Table
5. Previous events are recorded in the event log, which can be
Table 5 Event list
Description
The sensor is ready for calibration.
The current calibration is good.
The stabilization time during calibration expired.
No buffer is detected.
Resolution
Calibrate the sensor.
Do the steps in
Clean the sensor
and
Replace the salt bridge
do the calibration again.
Repeat the calibration with fresh reference
solutions.
Do the steps in
Clean the sensor
Then, do the calibration again.
Make sure that the correct temperature
element is selected. Refer to the TEMP
ELEMENT option in
Configure the sensor
on page 13.
Calibrate the sensor.
Note: To change the calibration interval, refer to the
CAL DAYS option in
Configure the sensor
on page 13.
Replace the sensor if necessary.
Note: To change the sensor replacement interval,
refer to the SENSOR DAYS option in
sensor
on page 13.
Calibrate the sensor.
Contact technical support.
The sensor is in air. Return the sensor to the
process.
The sensor is damaged or dirty. Do the steps
in
Clean the sensor
on page 18.
The standard cell solution leaked or
evaporated. Replace the standard cell
solution.
The reference electrode is damaged. Replace
the sensor.
Return to calibration.
on page 18
on page 18. Then,
on page 18.
Configure the
English 23