Startup instructions

Only qualified personnel should conduct the tasks described in this
section of the manual.
Setup the instrument in a properly ventilated place. The power supply
connector on the rear panel must remain accessible and close to the
user (2 m maximum) so that you can quickly disconnect the cables in
case of emergency.
The room temperature must be between 5 and 40 °C.
The relative humidity must be between 20 and 80%.
Use only a grounded socket for the connection of this device to the
power supply. If you are not sure if the sockets are grounded, have
this checked by a qualified electrician. The power plug serves in
addition to the power supply to isolate the device quickly from the
power source where necessary. During the disconnection from the
power source it must be made sure that the correct power plug is
pulled (for example by labelling the sockets). This is recommended
for long-term storage and can prevent potential dangers in the event
of a fault. Therefore make sure that the socket to which the device is
connected is easy to reach by each user at all times.
Electrical dangers and fire hazard. Use only the supplied power cable.
Connect the mains socket of the device to the mains supply using
the 3-lead power cord provided. The device must be connected to an
earthed mains socket for safety reasons. Efficient grounding is essential
for reliable measurements and security.
In the USA or Canada, use a UL listed power cord only.
Switch on the device (O/I power switch set to "I"). The device displays
an identification screen (name and embedded software version)
followed by the Main window.