Run sequence

Refer to Short-Form Reminder no. 7.
1.
Install the electrode then the reagent groups to be used to perform
the analysis.
2.
Display the Main window. Press 3 then 1 and prepare the sample
stack.
3.
Display the Main window. Press 1 to start the analysis.

Run direct measurement

Refer to Short-Form Reminder no. 8.
1.
Select the method.
2.
Install the electrode group.
Display the Electrodes window.
3.
4.
Press 2 to start the measurement.
5.
Select the electrode in the list. You can stop or start stirring by
pressing 1.
Running analyses