Service Instructions for Multichannel Pipette Tip Cones
To ensure even performance between all channels in a multichannel pipette, all tip cones have
to be changed at the same time, if any of them needs to be changed. Don't mix tip cones of
different packages, because one bag contains a matched set of tip cones.
Sterilization
The entire pipette can be sterilized by autoclaving it at 121°C (252°F) (2 ata) (minimum 20
minutes). No special preparations are needed for autoclaving. You can use steam sterilization
bags if needed.
After autoclaving the pipette must be cooled to room temperature for at least two hours. Before
pipetting, make sure that the pipette is dry. We recommend that you check the calibration after
every sterilization cycle to achieve the best possible accuracy.
The table below lists possible problems and their solutions.
Defect
Possible reason
Leakage
Tip incorrectly attached
Foreign particlesbetween tip and
tip cone
Foreign particles between the
piston, the O-ring and the cylinder
Insufficient amount of grease on
cylinder and O-ring
O-ring damaged
Incorrect operation
Inaccurate
dispensing
Tip incorrectly attached
Calibration altered: caused by
misuse, for example
Unsuitable calibration
Inaccurate
High viscosity liquids may require
dispensing with
recalibration
certain liquids
The Finnpipette F2 is shipped in a specially designed package containing the following items:
1.
The Finnpipette
2.
Service tool
3.
Multichannel service tool
4.
Finntip sample
5.
Tube of grease
6.
Instruction manual
7.
Calibration certificate
8.
Shelf hanger
9.
Two stickers
CAUTION!
The Finnpipette is designed to allow easy in-lab service. If you would prefer to
have us or your local representative service your pipette, please make sure that
the pipette has been decontaminated before you send it to us.
Please note that the postal authorities in your country may prohibit or restrict the
shipment of contaminated material by mail.
Trouble shooting
Package
11
Solution
Attach firmly
Clean tip cones attach new tips
Clean and grease O-ring and
cylinder.
Grease accordingly
Change the O-ring
Follow instructions carefully
Attach firmly
Recalibrate according to
instructions
Recalibrate with the liquids
in question