Using Your Column
Installation
The direction of flow is marked on the column.
For removable, zero-dead volume column connections, Agilent recommends
the use of the InfinityLab Quick Connect and Quick Turn family. Choices are:
Maximum
System Pressure
Up to 400 bar
Up to 800 bar
Up to 1300 bar
For more information and part numbers, please see to the Agilent InfinityLab Fitting
Brochure (5991-5164EN).
InfinityLab Quick Connect
assembly, p/n 5067-5961
Learn more at www.agilent.com/chem/infinitylabfittings
Column conditioning
Every column is tested before shipment and shipped in solvent, typically
methanol. Care should be taken to not exceed pressure limits when
transitioning mobile phase compositions. Flushing with 10 to 20 column
volumes should help in transitioning to your mobile phase. The column
should be properly equilibrated prior to use. This will ensure reproducibility
and help prevent retention time drifting.
Recommended Fitting
InfinityLab Quick Turn Fitting
(finger-tight)
InfinityLab Quick Turn Fitting
(with Mounting tool)
InfinityLab Quick Connect Fitting Fitting: 5067-5965
Part Numbers
Fitting: 5067-5966
Fitting: 5067-5966
Mounting Tool: 5043-0915
InfinityLab Quick Turn
fitting, p/n 5067-5966