10.4
Pressure Monitoring Test
1. Select the Standard Insufflation Mode as the operating mode.
2. Select a nominal pressure of 15 mmHg and adjust in the user menu the gas
flow rate 1.
3. Use the syringe to generate a pressure of at least 19 mmHg, which registers
on the manometer.
An acoustic warning sound is emitted with a pressure of more than 19 mmHg
(for 5 seconds) and the display depicts Overpressure.
4. Reduce the pressure.
The warning ends when the pressure falls below 19 mmHg (nominal pressure
plus 4 mmHg).
10.5
Max. device pressure test
(1)
(3)
1. Select the Standard Insufflation Mode as the operating mode.
2. Select the max. nominal gas flow.
3. Connect a manometer (1) and an open Veress cannula (3) to the insufflation
tube connection (2).
4. Select the max. gas flow.
5. Start insufflation:
Press the START key.
The manometer registers a pulsing pressure increase. When the pressure sta-
bilizes, the manometer indicates a maximum pressure between 65 and 75
mmHg.
6. Stop insufflation:
Press the STOP key.
10.6
Gas Flow Rate Test
Test setup with open connection, without connected insufflation tube.
1. Adjust in the user menu the gas flow rate 2 (20 liters).
2. Start insufflation: Press the START key.
3. Press the RESET key (0.0 l must be depicted).
Now start the measuring for one minute.
4. Stop the insufflation after one minute. Press the STOP key.
The gas consumption should be at least 17.5 - 22.5 liters.
(2)
Annual Inspection
EN
35