Annual Inspection
EN
34
6. Select the max. nominal gas flow.
The following values are displayed:
Actual gas flow equals nominal gas flow rate
Actual pressure 0 mmHg
Initial Insufflation is displayed.
Streaming gas can be heard at the insufflation tube connection (barbed con-
nector).
7. Stop insufflation:
Press the STOP key.
The following values are displayed:
Actual pressure 0 mmHg
Gas consumption > 0.0 l
8. Press the RESET key.
Gas consumption 0.0 l
The basic function check of the device is complete.
10.3

Pressure Sensor Test

(1)
(3)
1. Select the Standard Insufflation Mode as the operating mode.
2. Do not press the START/STOP key.
CAUTION!
Never use the syringe to extract gas from the device.
3. Connect a manometer (1) and an air-filled syringe (3) to the insufflation
tube connection (2).
4. Use the syringe to generate a pressure of at least 10 mmHg, which is indicat-
ed on the manometer.
Actual pressure display: 10 ±2 [mmHg]
5. Use the syringe to generate a pressure of at least 20 mmHg, which is indicat-
ed on the manometer.
Actual pressure display: 20 ±2 [mmHg]
6. Use the syringe to generate a pressure of at least 30 mmHg, which registers
on the manometer.
Actual pressure display: 30 ±2 [mmHg]
(2)