Follow on screen messages.
9. When procedure is completed stop insufflation by pressing the STOP key.
10. When working in the AirSeal Mode, it is possible that body fluids can be ex-
tracted from the surgical field over the evacuation line and into the filter
housing. In order to prevent contamination of the device, bodily fluid is re-
tained by the fluid trap of the housing component. The capacity of the fluid
trap is limited. A warning message will be emitted if the fill level LOW is
reached. Cannula and tubing placement should be checked to make sure no
more fluid enters the filter. Insufflation can be continued with full function-
ality. If fluid level HIGH is reached, the AirSeal function will shut down. Insert
obturator in AirSeal Cannula to maintain pressure with normal insufflation.
Insert new tri-lumen tube to continue with AirSeal Insufflation.