7. Maintenance
ErgoOne® is a low maintenance pipette. However, individual working
methods, length of application and internal quality control guidelines
mean that regular monitoring and maintenance of your pipette is
necessary.
Note:
If you make any changes to the pipette or replace individual
parts, a check with recalibration is required. If the pipette cannot be
calibrated successfully or if a fault continues, contact STARLAB.
7.1 Maintaining
Single-Channel Pipettes
The figures shown below refer to the 200µl model. They may differ
slightly for other pipette models.The maintenance work is carried out
with the aid of the auxiliary tool supplied (tool not required for 5ml
pipette) (Fig. 3)
A
F or removing and replacing the seals
A
on the 100µl and 1000µl models
F or removing and replacing the seals
B
on the 200µl model
F or loosening the lower housing of
C
pipettes up to the 1000µl model
Disassembly and cleaning up
to the 1000µl model
•
Fully depress the tip ejector
button
and pull the tip ejector
2
down (Fig. 4).
5
•
Carefully unscrew the tip holder
by turning the auxiliary tool
7
clockwise (ErgoOne® 5ml tip
holder can be removed without
the auxiliary tool).
C
12
ErgoOne
®
B
Fig. 3
User Manual
Fig. 4
15