Measurement range setting in USgal/min (S4 off)
6.3
If a display is connected to the converter electronics, then the settings of the DIP switch which are used
to set the equipment without a display are ignored. The converter can now be set using a simple, opera-
tor-guided menu with plain language in the display.
The advantages of this version over the version without a display are:
1.
2.
3.
4.
Measurement range table
Display format
The display in the Miniflow is a graphic-compatible display in 97*32 point format.

Process display

The first line of the process display indicates the value of the present flow and the second line indicates
the unit.
The counter status is shown in the lower line litres or US gallons. During the alarm or other fault reporting,
the display disappears and the fault reporting appears in plain language.
GB 18
FXL5000 (Miniflow)
DN
Stage 1
S1
S2
S3
Off
Off
Off
10
1.0 gal/min
15
1.5 gal/min
25
4.0 gal/min
50
16.0 gal/min
DN
Stage 1
S1
S2
S3
On
Off
Off
10
3.0 gal/min
15
7.0 gal/min
25
20.0 gal/min
50
80.0 gal/min

Setting Miniflow, version with display

Measurement range and damping can now be set in infinitely variable manner
Operator is free to choose from several physical units for flow
The optocoupler output can – when configured as contact output
assume several functions. (e.g. min. alarm, max. alarm etc.)
The pulse value can now be set!
DN Measurement range can be set (in infinitely variable manner) in l/min between
10
min. 0 -
2.5 l/min
15
min. 0 -
5
l/min
25
min. 0 -
15
l/min
50
min. 0 -
60
l/min
DN Measurement range can be set (in infinitely variable manner) in USgal/min between
10
min. 0 -
1.0 gal/min
15
min. 0 -
1.5 gal/min
25
min. 0 -
4
gal/min
50
min. 0 -
16
gal/min
>V
122.5
l/min
>V
3256 l
Stage 2
S1
S2
S3
S1
Off
Off
On
Off
1.5 gal/min
2.0 gal/min
2.5 gal/min
4.0 gal/min
8.0 gal/min
12.0 gal/min
30.0 gal/min
50.0 gal/min
Stage 2
S1
S2
S3
S1
On
Off
On
On
4.0 gal/min
5.0 gal/min
8.0 gal/min
10.0 gal/min
25.0 gal/min
40.0 gal/min
100.0 gal/min
160.0 gal/min
max. 0 -
50
max. 0 - 100
max. 0 - 300
max. 0 - 1200
max. 0 -
10
max. 0 -
25
max. 0 -
80
max. 0 - 320
Stage 3
Stage 4
S2
S3
S1
S2
On
Off
Off
On
2.5 gal/min
5.0 gal/min
15.0 gal/min
60.0 gal/min
Stage 3
Stage 4
S2
S3
S1
S2
On
Off
On
On
10.0 gal/min
25.0 gal/min
80.0 gal/min
320.0 gal/min
l/min
l/min
l/min
l/min
gal/min
gal/min
gal/min
gal/min
6 Operation
S3
On
S3
On
D184B112U04