... the display shows
– The power supply from the mains was interrupted during measuring. If the rechargea-
ble battery is fully charged, measuring continues for at least 4 hours. Restore the pow-
er supply as soon as possible.
bAtt
...
appears in the display?
– The rechargeable batteries have run down so far that measuring is no longer possible
without a power supply. Restore the power supply.
– If this fault occurs quite frequently, have the rechargeable batteries checked.
St0p
...
appears in the display?
– Maximum load has been exceeded.
... the display flashes?
– If you have just activated the Hold or Tare function, please wait until the scale has de-
tected a steady weight value. The display will then stop flashing.
– If there has been no activation of a function beforehand, remove the load from the scale
0. 0
and wait until
... the display
– The ambient temperature of the scale is too high or too low. Place the scale in an
ambient temperature between +10 °C and +40 °C. Wait about 15 minutes for the
scale to adapt to the ambient temperature and then weigh again.
Er:1:11
... the display
– At least one weighing cell is not correctly connected or the connection is back to front.
Check the connections. The colours of plug and socket must match.
Switch the scale off using the Start key and restart. The scale then works normally
again. If this is not the case, inform Servicing.
Er:1:12
... the display
– The maximum weight for switching on was exceeded when the scale was switched
on, e.g. if bed and patient are being weighed. Remove the load from the scale and start
it again. The scale then works normally again. If this is not the case, inform Servicing.
... one segment is permanently illuminated or is not illuminated at all?
– The relevant segment is faulty. Inform Servicing.
30
?
is displayed, then weigh again.
appears?
appears?
appears?