E N G L I S H

IEC SYMBOLS

Attention, consult instruction guide
AC current
Electric Shock Hazard. Do Not Open

IMPORTANT SAFEGUARDS

When using electrical products, especially when children are present, basic safety precautions should always be followed. Read
all instructions before using. Important information is highlighted by these terms:
DANGER
Urgent safety information for hazards that will cause serious injury or death.
WARNING
Important safety information for hazards that might cause serious injury.
CAUTION
Information for preventing damage to the product.
NOTE
Information to which you should pay special attention.
Important safeguards are indicated throughout this guide; pay special attention to all safety information.
READ ALL INSTRUCTIONS BEFORE USING.
SAVE THESE INSTRUCTIONS.
GENERAL DANGERS & WARNINGS
In order to ensure the safe installation, assembly, and operation of the personal oxygen station, these instructions MUST be followed:
DANGER-NO SMOKING
DANGER
DANGER-Electric Shock Hazard. Do not disassemble. The DeVilbiss iFill Personal Oxygen Station contains no
user serviceable parts. If service is required, contact your Sunrise Medical provider or authorized service center.
DANGER-Fire Hazard. Do not lubricate. Do not allow grease or oil from your hands or other source to come into
contact with the regulator or cylinder valve connection. These solutions may be flammable and cause injury.
DANGER-Keep the personal oxygen station at least 5 feet (1.6 m) from hot, sparking objects or naked sources of
flame. Position your unit at least 6 inches (16 cm) from walls, draperies, or any other object that might prevent the
proper flow of air in and out of your oxygen station. The oxygen station should be located so as to avoid pollutants
or fumes.
DANGER-Oxygen causes rapid burning. Do not smoke while your personal oxygen station is operating, or when
you are near a person utilizing oxygen therapy. Keep the oxygen station at least 5 feet (1.6 m) from hot, sparking
objects or naked sources of flame.
DANGER-Do not lay the cannula down while the cylinder is delivering oxygen. High concentrations of oxygen can
cause rapid burning.
WARNING
• Do not install, assemble, or operate this equipment without first reading and understanding this instruction man-
ual and the oxygen cylinder warning label. If you are unable to understand the warnings, cautions and instruc-
tions, contact your provider or technical personnel if applicable before attempting to install or use this equip-
ment - otherwise, injury or damage may occur.
• The use of this device is limited to an oxygen patient. Cylinders MUST be used ONLY by an oxygen patient and
are not to be distributed to any other individual for any purpose.
• Do not use parts, accessories, or adapters other than those authorized by Sunrise Medical.
• NEVER block the air openings of the product or place it on a soft surface, such as a bed or couch, where the air
opening may be blocked. Keep the openings free from lint, hair and the like.
• If the oxygen station has a damaged cord or plug, if it is not working properly, if it has been dropped or dam-
aged, or if it has been submersed in water, do not use and call a qualified technician for examination and repair.
A-535
Type BF applied part
Double Insulated
Danger-No Smoking
4
4
I M P O R TAN T SA FE G UA R D S