Koldt loddestedskomp. (CJC) ..................... < ±1,0°C
Følerfejlsdetektering .................................... Ja
Følerfejlsstrøm:
under detektering ................................. nom. 33 mA
ellers ..................................................... 0 mA
Spændingsindgange:
Måleområde ............................................... -12...150 mV
Min. måleområde (span) .............................. 5 mV
Indgangsmodstand ..................................... 10 MΩ
Udgange:
Strømudgange:
Signalområde .............................................. 4...20 mA
Min. signalområde ....................................... 16 mA
Opdateringstid ............................................. 440 ms
Udgangssignal ved EEpromfejl ................... ≤ 3,5 mA
Belastningsmodstand .................................. ≤ (V
Belastningsstabilitet .......................................... < ±0,01% af span / 100 Ω
Følerfejlsdetektering:
Programmerbar ........................................... 3,5...23 mA
NAMUR NE43 Upscale ............................... 23 mA
NAMUR NE43 Downscale ........................... 3,5 mA
Ex- / I.S.-data:
Udgang, klemme 11...13, (21...23):
U i ................................................................. : 28 VDC
I i .................................................................. : 120 mADC
P i ................................................................. : 0,84 W
L i ................................................................. : 10 µH
C i ................................................................. : 1,0 nF
Indgang, klemme 41...44, (51...54):
U o ............................................................... : 9,6 VDC
I o ................................................................. : 25,1 mADC
P o ................................................................ : 60,3 mW
L o ................................................................ : 45 mH
C o ................................................................ : 3,5 µF
EEx- / I.S.-godkendelse:
DEMKO 99ATEX126960 ..............................
Max. omgivelsestemp. for T1...T6 .............. 60°C
Anvendes i zone .......................................... 0, 1 eller 2
8
All manuals and user guides at all-guides.com
- 7,2) / 0,023 [Ω]
forsyn.
II 1 G
EEx ia IIC T1...T6
GOST R godkendelse:
Standard: VNIIM .......................................... Cert. no. Ross DK.ME48.V01899
Ex: VNIIFTRI ................................................ Cert. no. Ross DK.GB06.V00100
Overholdte myndighedskrav:
EMC 2004/108/EF
Emission og immunitet ........................ EN 61326
ATEX 94/9/EF .............................................. EN 50014 og EN 50020
Af span = Af det aktuelt valgte område
Standard: