Instant Help - KERN and SOHN SFB 10K1HIP Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

11 Instant help

In case of an error in the program process, briefly turn off the balance and
disconnect from power supply. The weighing process must then be restarted from
the beginning.
Help:
Fault
The displayed weight does
not glow.
The displayed weight is
permanently changing
The weighing result is
obviously incorrect
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.
52
Possible cause
• The balance is not switched on.
• Mains power failure (mains cable defective).
• Power supply interrupted.
• (Rechargeable) batteries are inserted incorrectly or
empty
• No (rechargeable) batteries inserted.
• Draught/air movement
• Table/floor vibrations
• Weighing plate has contact with other objects.
• Electromagnetic fields / static charging (choose
different location/switch off interfering device if
possible)
• The display of the balance is not at zero
• Adjustment is no longer correct.
• Great fluctuations in temperature.
• Warm-up time was ignored.
• Electromagnetic fields / static charging (choose
different location/switch off interfering device if
possible)
SFB-BA-e-1625

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Sfb 15k5hipmSfb 20k2hip

Table des Matières