KERN and SOHN SFB 10K1HIP Mode D'emploi page 90

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

 Press
; weighing system is in tolerance weighing mode.
From here evaluation takes place whether the goods to be
weighed are within the two tolerance limits.
Weighing with tolerance range
 Set item weight, see chap. 7.10.
 Tare when using a weighing container.
 Put on goods to be weighed, tolerance control is started. The signal lights
indicate whether the load is within the two set limits.
Load below specified
tolerance
Red signal light
next to „-" ON illuminated
• The tolerance control is not active when the weight is under 20d.
• To delete limits, enter „00000 PCS".
34
Load within specified
tolerance
Green signal light
next to „„ illuminated
Load exceeds specified
tolerance
Red signal light
next to „+" ON illuminated
SFB-BA-e-1625

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Sfb 15k5hipmSfb 20k2hip

Table des Matières