In case of an error in the program process, briefly turn off the display unit and
disconnect from power supply. The weighing process must then be restarted from the
beginning.
Help:
Fault
The displayed weight does
not glow.
The displayed weight is
permanently changing
The weighing result is
obviously incorrect
76
Possible cause
• The display unit is not switched on.
• Mains power supply interrupted (mains cable
defective).
• Power supply interrupted.
• (Rechargeable) batteries are inserted incorrectly or
empty
• No (rechargeable) batteries inserted.
• Draught/air movement
• Table/floor vibrations
• Weighing pan has contact with other objects.
• Electromagnetic fields / static charging (choose
different location/switch off interfering device if
possible)
• The display of the balance is not at zero
• Adjustment is no longer correct.
• Great fluctuations in temperature.
• Warm-up time was ignored.
• Electromagnetic fields / static charging (choose
different location/switch off interfering device if
possible)
KXS/KXG-TM-BA_IA-e-1621