8.5 Instant help

In case of an error in the program process, briefly turn off the balance and disconnect
from power supply. The weighing process must then be restarted from the beginning.
Help:
Fault
The displayed weight is permanently
changing
The weighing result is obviously
incorrect
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.
KFP V20NM-IA-e-1712
Possible cause
• Draught/air movement
• Floor vibrations
• Weighing plate has contact with other
objects.
• Electromagnetic fields / static charging
(choose different location/switch off
interfering device if possible)
• No zero display with unloaded balance
• Adjustment is no longer correct.
• Great fluctuations in temperature.
• Weighing bridge on an uneven surface.
• Electromagnetic fields / static charging
(choose different location/switch off
interfering device if possible)
13