System settings

Fig. 236
From the main menu
(Fig.
236) you can access the system settings with
<SYS> (Fig. 237).
Fig. 237
Setting the national language was already described in  2.5.
6.1

Calibration settings

The Calibration settings item is used to select the buffers for the calibration of the pH electrode as well as to set
the temperature of the buffer solution (Fig. 238).
The temperature has only to be set if neither a resistance thermometer (Pt 1000), nor a pH electrode with an
integrated temperature measurement probe is connected.
Fig. 238