Technical Specifications; Titrator Titroline 7800 - Xylem TitroLine 7800 Mode D'emploi

Titrateur
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

150
1.3

Technical Specifications

Titrator TitroLine
1.3.1
Translation of the legally binding German version
EMC compatibility according to the Council Directive: 2014/30/EU;
applied harmonized standards: EN 61326-1: 2013
Low-voltage directive according to the Council Directive 2014/35/EU;
Testing basis EN 61 010-1: 2010 for laboratory equipment
RoHS Council Directive 2011/65/EU
FCC Part 15B and ICES 003
Country of origin: Germany, Made in Germany
The following solvents/titration reagents are allowed to be used:
 All common titration solutions.
 As reagent water and all non-aggressive non-organic and organic fluids are allowed.
 If using combustible fluids fire please adhere to the Guidelines for Explosion Protection and Prevention of the
chemical industry.
 For fluids with higher viscosity ( 5 mm
speed can be adjusted.
 Fluids with viscosity over 20mm
To ensure maximum accuracy of the readings we recommend to allow some reasonable time for the
®
7800 to "warm up".
TitroLine
Measuring input 1 (analog):
pH/mV-input with 24 bit transducer for high-precision readings.
Electrode socket according to DIN 19 262 or additional with BNC socket insert (Z 860).
Reference electrode 1 x 4 mm socket.
Adjustable damping settings of the pH/mV measuring signal.
RFID receiver for SI Analytics
Measurement range
pH
pH
- 3.0 ... 18.00
U mV
mV
- 2000 ... 2000
Measuring input (Pt 1000):
Temperature sensor - connector for a resistance thermometer Pt 1000
and NTC 30 kOhm. Connection: 2 x 4 mm sockets.
Measuring input (μA):
Karl Fischer (Dead-Stop) connector (µA) for double platinum electrode.
Polarisation voltage variably adjustable from 40 ... 220 mV.
Connector: 2 x 4 mm – sockets.
* The measurement uncertainty of the sensor probe has to be taken into account as well.
®
7800
2
/s), lower boiling point or affinity to outgas, the filling and dosage
2
/s cannot be dosed.
Over range
- 3.1 ... 18.00
- 2020 ... 2020
Measurement
range T [°C]
Pt 1000
- 75 ... 195
- 40 ... 0
NTC 30
0 ... 125
Measurement
range I [A]
100
50
10
5
®
ID electrodes.
Measurement accuracy*
Display resolution
0.001
0.1
Measurement accuracy* without
Display resolution
0.1
0.1
0.1
Measurement accuracy* without
Display resolution
0.1
0.1
0.1
0.1
(Release: 16. August 2018)
without sensor probe
 1 Digit
0.002
 1 Digit
0.10
sensor probe
0.2 K  1 Digit
1.0 K  1 Digit
0.3 K  1 Digit
sensor probe
-5 /+ 3 A  1 Digit
+/- 3 A  1 Digit
+/- 1 A  1 Digit
+/- 0.2 A  1 Digit
Input
resistance 
13
> 1  10
13
> 1  10

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Si analytics titroline 7000

Table des Matières